Model: 905V
Hãng: Quantek Instruments - Mỹ
Sử dụng để phân tích khí oxy trong các sản phẩm dược phẩm - lọ, vỉ thuốc, hoặc chai. Đạt được kết quả với ít nhất là 0.5cc các khoảng trống.
Thông số kỹ thuật
Range: 0.0 to 100% Oxygen
Resolution: 0.01% Oxygen
Accuracy: +/- 0.1% (below oxygen levels of 10%)
Minimum Detection Limit: 0.01%
Drift: less than 1% per month
Sensor: Proprietary electrochemical cell, expected lifetime 4-5 years
Sample Method: Water displacement method; alternately, product can be squeezed to introduce sample to analyzer
Minimum Sample Required: 0.5cc
Sampling Assembly: Probe with male luer connection for side-port needle, PTFE tubing
Calibration: Calibrates with room air; coarse SPAN adjustment on back; fine span adjustment on side; ZERO adjustment on back
Calibration Frequency: Minimum once per year
Power Supply: 100-240V (50-60Hz) Charger included, with US or international interchangeable plug adapter (upon request)
Backup Power: Optional 12V internal battery pack
Input Voltage: 100-240V, 50-60Hz
(Optional) Battery Life: 10-14 hours before recharging
Size: 9.8in x 4in x 10in. (249 x 102 x 254 mm)
Weight: 8 lb. (3.6 Kg)
Warranty: 2 years, parts and labor
Standards: CE, RoHS
Included Consumables: Low volume sample probe, three zero volume puncture needles, tubing & tee fitting, two plastic 1cc and 5cc syringes, 115/240Vac to 12Vdc power adapter
Included Documentation: Owners manual (includes operating instructions with detailed calibration instructions, calibration certficate, precautions), NIST calibration certificate
Origin of Goods: Our products are manufactured in the U.S.A.
Pharmaceutical vials
Blister Packs
Cosmetic Products
Bottles with limited Headspace
(with container test kit) - Ampoules
(with container test kit) - blister packs